On-cruise calibration of XBT probes


Contact
gbudeus [ at ] awi-bremerhaven.de

Abstract

A procedure is described to calibrate XBT probes at one temperature point immediately before launching. The individual XBT temperature offsets are determined by using a stirred calibration bath and a stable, accurate reference thermometer designed for field use. The proposed calibration procedure increases XBT temperature accuracy considerably, while it hardly raises time and effort spent on standard XBT drops. The performance of the system in the field and in the laboratory is described. It is shown that the achieved temperature accuracy is about 0.03 K, and is basically limited by the used reference thermometer. © 1993.



Item Type
Article
Authors
Divisions
Programs
Publication Status
Published
Eprint ID
497
DOI https://www.doi.org/10.1016/0967-0637(93)90116-k

Cite as
Budéus, G. and Krause, G. (1993): On-cruise calibration of XBT probes , Deep Sea Research Part I: Oceanographic Research Papers, 40 (7), pp. 1359-1363 . doi: https://www.doi.org/10.1016/0967-0637(93)90116-k